Dynatech MP-7000 Manual de Serviço Página 66

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 65
62 9650-0801-01 Rev. A
14 Tilt side panel to access the EtCO
2
connector and use an orange stick release
the tabs and remove the EtCO
2
connector.
15 Remove the EtCO2 module from the unit.
16 Remove three (3) screws from the outside
of the left side panel in the areas illustrated.
Note: During reassembly, torque screws to
6 in-lbs.
17 Remove four (4) screws from the inside top
of the left side panel in the areas illustrated
and carefully pull out side panel.
Note: During reassembly, torque screws to
6 in-lbs.
18 Remove ECG ZIF connector in the area
illustrated.
19 Remove the left side panel from the unit.
EtCo
2
connector
before removal
EtCo
2
connector
after removal
EtCO2 connector after removal
Vista de página 65
1 2 ... 61 62 63 64 65 66 67 68 69 70 71 ... 91 92

Comentários a estes Manuais

Sem comentários