9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 69
Chapter 4
Life Cycle Counters
Overview
This chapter contains instructions for setting life cycle counters on the Propaq MD. The Propaq
MD tracks various life cycle counters for specific modules installed in the device. When
replacing modules in the unit, it is important to reset the life cycle counters. Once you have
reset the appropriate life cycle counter(s), see “Functional Verification Tests” on page 1-21 to
perform the functional verification tests to ensure that the functions of the Propaq MD unit
work properly.
Note: Only reset the life cycle counters for the modules that you have replaced.
Resetting Life Cycle Counters
NIBP
On the Propaq MD, do the following:
1. Press the Setup quick access key ( ).
2. Use the navigation keys to select Supervisor>Service>Diagnostic>NIBP>Cycle/Error
Counts.
3. Select Reset Counters. Select Yes when prompted to confirm the reset.
See “Functional Verification Tests” on page 1-21 to perform the functional verification tests.