15
Electrosurgery Analyzers
Key features
Easy to use simple
configuration
Oscilloscope output, high-
frequency leakage, and return
electrode contact quality moni-
tor tests
Instantaneous and signal-
averaging measurement mode
Ability to connect with Fluke
Biomedical’s medTester 5000C
for automated solution
RS-232 port for computer
control
Battery powered
4-digit numeric LCD with back-
light and power-save mode
RF303
RS
Electrosurgery
Analyzer provides enough
user-selectable test loads
to do routine maintenance
checks on most electrosur-
gery units on the market
today. Compact and por-
table, the device is so
simple to use that techni-
cians can become proficient
with the RF303
RS
within
minutes.
The unit measures ESU
output and high-frequency
leakage, allows for verifi-
cation tests on the return electrode contact quality monitors, and has an
oscilloscope output for waveform viewing. Instantaneous output or select-
able sample times provide extra versatility. The instantaneous mode is
sufficient for most units, but if output readings are variable and require
stabilizing, the signal-averaging mode allows users to manually select two
additional, slower sampling times to produce an accurate average reading.
Modes of operation Line powered, battery powered, offline (battery maintenance
charge)
Test parameters
Power (W), HF current (mA), test load (Ω)
Tests performed
generator output
HF leakage (performs to IEC 601 2-2, 1289-2, ANSI/AAMI standards)
Type BF test 1: Earth-referenced monopolar output
Type BF test 2: Earth-referenced monopolar output
Type CF/bipolar: Isolated monopolar or bipolar output
Current measurement
(leakage)
Range: 30 mA to 2500 mA rms
Resolution: 1 mA
Accuracy: ± 2.5 % of reading or ± 15 mA (whichever is greater)
Power measurement
(output)
Range: 1 W to 400 W
Resolution: 0.1 W
Accuracy: ± 5 % of reading or ± 3 W (whichever is greater)
Bandwidth of rms
converter circuit
(1 % accuracy)
Flat response: 10 kHz to 10 MHz
-3 dB points: 1 kHz to 20 MHz
Frequency response
System response: -3 dB points, 1 kHz to 10 MHz at 300 Ω
RECQM test
50 Ω to 750 Ω, 50 Ω steps
Test load section Number of selections: 15
Range: 50 Ω to 750 Ω
Step size: 50 Ω
Accuracy (dc to 500 KHz): ± 4 % of selected value measured at
calibration to ± 1 % (across the entire operating temperature range)
Duty cycle: 25 % @ 400 W (max 30 sec ON during any 2 minute period)
Resonance impedance variation: ± 0.5 dB max (< 10 MHz)
Auxiliary leakage test
load
Fixed: 200 Ω
Accuracy: ± 4 %
Power rating: 225 W
Input capacitance
(nominal)
Active to dispersive: 30 pF
Active or dispersive to earth ground: 40 pF
Oscilloscope output Transformer coupled output, uncalibrated
Connector type: BNC
RF303
RS
Electrosurgery Analyzer
Specications