9
Key features
• Defibrillator analyzer
• Monophasic and biphasic dc
energy measurement
• Energy and cardioversion mea-
surement
• Peak voltage, peak current, and
overshoot measurement
• 2-line x 24-character display
• Bidirectional RS-232 port for
computer control
• Storage and playback of output
waveform so results can be
viewed in greater detail
• Optional Ansur test automation
software to standardize testing
procedures, capture waveforms
and test results, and print and
document test results
The QED 6 provides a
scalable solution to accu-
rately test defibrillators.
Lightweight and portable,
the QED 6 measures a
wide range of defibrillator
energy output parameters.
An RS-232 serial
port supports remote
computer control and test
documentation.
Output energy test
Load 50 ± 1 %, with inductance < 70 µ h
Resolution High-range: 1 J
Low-range: 0.1 J
Low-range 0 J to 100 J
High-range 0 J to 400 J
Pulse width 1 ms to 50 ms
Maximum current Low: 35 A
High: 110 A
Maximum voltage Low: 1750 V
High: 5500 V
Minimum voltage Low: 20 V
High: 66 V
Accuracy 400 J range: ± 2 % of reading
100 J to 400 J: ± 2 J
100 J Range: ± 2 % of reading, ± 0.1 J
Waveform storage Discharge viewable via ECG output, paddles, and scope
output
Time expansion lead II
amplitude
High = 3000 V / mV
Low = 900 V / mV
Peak/overshoot
Voltage accuracy 400 J range: ± 10 V
100 J Range: ± 25 V
Current accuracy: ± 1 A
Cardioversion synchronization
test
Measurement from peak or base of simulated R-wave: 0
ms to 199.9 ms
Accuracy: 1 % of full scale or ± 2 ms, whichever is
greater
QED 6
Debrillator Analyzer
Specications
Defibrillator/Transcutaneous Pacer Analyzers