36 Multi-Function Simulators
Oximeter SpO
2
optical emitter and detector (optional)
% O
2
Range 30 % to 100 %
Resolution 1 %
% O
2
accuracy With oximeter manufacturer’s R-curve Saturation within UUT specific range: ± (1 count +
specified accuracy of the UUT)
Saturation outside UUT specific range: monotonic with
unspecified accuracy
With Fluke Biomedical R-curves 91 % to 100 % ± (3 counts + specified accuracy of
the UUT)
81 % to 90 % ± (5 counts + specified accuracy of the
UUT)
71 % to 80 % ± (7 counts + specified accuracy of the
UUT)
Below 71 % monotonic with unspecified accuracy
Heart rate 30 BPM to 300 BPM in 1 BPM steps. Oximeter SpO
2
optical emitter and detector is synchronized with ECG rate
delayed by 150 ms.
Transmission: ratio of detector
current to LED current, expressed
in parts per million (ppm)
Range 0 ppm to 300.00 ppm
Resolution 0.01 ppm
Accuracy + 50 %/- 30 % for compatible monitors, unspecified
for others. Selected by finger size and color: dark,
thick finger, medium finger, light, thin finger, neonatal
foot.
Pulse amplitude Range 0 % to 20.00 %
Resolution 0.01 %
Artifact Respiration Range: 0 % to 5 % of transmission
Resolution: 1 %
Rate: all ProSim respiration simulation settings
Ambient light Range: 0 to 5X transmitted light
Resolution: 1X
Frequency: DC, 50 Hz, 60 Hz, and 1 kHz to 10 kHz in
1 kHz steps
Masimo Rainbow technology Masimo Rainbow technology with an optional adapter supplied by Masimo that allows the ProSim two
wavelength to test the Rainbow multiple wavelength system
Compatible manufacturer
products
With manufacturer R-curve Nellcor, Masimo, Nonin, and Nihon Kohden
With Fluke R-curve Mindray, GE-Ohmeda, Philips/HP, and BCI
ProSim 8
Vital Signs Simulator
Pre-Defined Simulations
Normal
Hypertensive
Hypotensive
Tachycardic
Bradycardic
Ventricular fibrillation
Asystole
Autosequences (default)
Monitor testing sequence
Medical training sequence
Oximeter testing sequence
Cardiac failure sequence
Arrhythmia sequence
Exercise sequence
Respiration sequence
Performance wave test
IBP testing sequence
Temperature sequence
Optional accessories
2392199 CI-3 Cardiac Output Box
3408564 Mini-DIN to DIN IBP Adapter
3890640 NIBP Test Chamber 500ML
4034627 Ansur Test Software ProSim
8 Plug-In
Cable kits
3984910 ProSim 8 Accessory Kit
(includes DIN to minDin adapter,
HP/Philips Intellivue IBP cable, GE
Marquette Eagle/Dash/Solar IBP
cable, Welch Allyn Propaq/SpaceLabs
Ultraview IBP cable, USB wireless
dongle, YSI400 series temperature
cable, YSI700 series temperature
cable, CI-3 Cardiac Output Box, spare
battery pack)
3984922 HP/Philips Intellivue Cable
Set
3984968 GE Marquette Eagle/Dash/
Solar Cable Set
3984946 ProSim 8 SpaceLabs
Ultraview Cable set
3984979 Welch Allyn/Propaq Cable
Set