42 Patient Simulators
Key features
12-lead ECG simulation with
independent outputs
43 arrhythmia selections
Four invasive BP channels,
including Swan-Ganz
simulation
Respiration and temperature
simulations
ECG performance testing,
including R-Wave detection
Large, bright 4-line x
20-character display
RS-232 serial port
Expansion port
Compact and lightweight
The MPS450 is Fluke Biomedi-
cal’s next-generation, portable,
multiparameter patient simu-
lator for your comprehensive
testing and training needs.
Whether its a quick check on
a bedside monitor, arrhythmia
recognition training, or per-
forming a complete PM on the
latest patient-monitoring sys-
tems, this simulator is a clear
choice with its broad range of
physiological waveforms, easy-
to-use interface, and compact,
portable size.
ECG normal sinus
rhythm
12-lead configuration with independent outputs
Amplitude: 0.05 mV to 5.5 mV
Rates: 30 BPM to 300 BPM
ECG waveform selections: Adult or pediatric
Superimposed artifact: 50 Hz and 60 Hz, muscle, baseline wander,
respiration
ECG performance Amplitude: 0.05 mV to 5.5 mV
Square wave: 2 Hz , 0.125 Hz
Pulse: 30 BPM, 60 BPM, 60 ms pulse width
Sine waves: 0.5 Hz to 100 Hz
Triangle wave: 2 Hz, 2.5 Hz
ST segments Elevated/depressed: -0.8 mV to 0.8 mV in 0.1 mV steps; plus
-0.05 mV and 0.05 mV steps
Accuracy All amplitudes ± 2 % of setting Lead II
All rates ± 1 %
All widths ± 1 %
Arrhythmia selections
(43 Total)
Premature rhythms
Supraventricular rhythms
Ventricular rhythms
Conduction defects
Pacemaker
Respiration
Baseline impedance: 500 Ω to 2000 Ω, leads I, II, III
Impedance variations: 3 Ω, 1 Ω, 0.5 Ω, 0.2 Ω
Rates: 15 BrPM to 120 BrPM and APNEA
Apnea periods: 12 seconds, 22 seconds, 32 seconds, and
continuous
Model MPS450 medSim 300B PS420 PS410 DataSim 6100
Arrhythmia selections 43 32 35 35 34
Respiration Yes Yes Yes No Yes
BP channels Yes, four Yes, four Yes, two No Yes, three
Swan-Ganz procedure Yes Yes Yes No Yes
Temperature channels Yes, one Yes, two Yes, one No No
User-programmable
auto-sequences
Yes (with
HHC3)
Yes, internal feature
plus HHC3 capability
No No Yes
Cardiac output Optional Optional Standard No Optional