Dynatech MP-7000 Especificações Página 55

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 68
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 54
55
Ordering information
VT PLUS HF Gas Flow Analyzer
VT+HF-US120 United States, 120 V
VT+HF-AUS250V Australia, 250 V
VT+HF-SHK250V Schuko, 250 V
VT+HF-UK250V United Kingdom, 250 V
VT+HF-BRAZ250 Brazil, 250V
Premium Precision Ventilator Test Kit
Includes: VT PLUS HF Gas Flow
Analyzer; and ACCU LUNG Portable
Precision Test Lung
VT+HG/ACCULUNG-US United States,
120 V
VT+HF/ACCULUNG-AUS Australia, 250 V
VT+HF/ACCULUNG-SHK Schuko, 250 V
VT+HF/ACCULUNG-UK United
Kingdom, 250 V
VT+HF/ACCULUNG-BRAZ Brazil, 250 V
VT-Plus Upgrades
(adds HF capability and RT-200 mode)
8831007 VT PLUS HF hardware and
firmware factory service upgrade (for
units lower than hardware v1.01.01;
additional flat-rate charge required for
factory service/calibration)
Included accessories
9VT0015 Operators Manual
8830200FG VT for Windows
®
PC
Software
75034 Serial Cable
1HD0011 Tilt Stand
Power cord (country specific)
VT-PLUS-7001 Accessory Kit (includes
16 accessories)
Ventilator parameter Resolution Range Accuracy
Inspiratory and expira-
tory tidal volume
0.1 ml As specified in high-flow/low-flow specification
Expiratory minute vol-
ume
0.001 lpm 0 L to 60 L ± 3 %
Breath rate 0.1 BPM 0.5 BPM to 150 BPM ± 1 %
Inspiratory-to expiratory
time ratio (I:E ratio)
0.01 1:200 to 200:1 ± 2 % or ± 0.1 s
Inspiratory time 0.01 s 0 s to 60 s ± 0.5 % or ± 0.02 s
Expiratory time 0.01 s 0 s to 90 s ± 0.5 % or ± 0.01 s
Peak inspiratory pressure 0.1 cmH
2
0 ± 120 cmH
2
O ± 3 % or ± 1 cmH
2
O
Inspiratory pause pressure 0.1 cmH
2
0 ± 120 cmH
2
O ± 3 % or ± 1 cmH
2
O
Mean airway pressure 0.1 cmH
2
0 ± 80 cmH
2
O ± 3 % or ± 0.5 cmH
2
O
Positive end-expiratory
pressure (PEEP)
0.1 cmH
2
0 -5 cmH
2
O to 40 cmH
2
O ± 3 % or ± 0.5 cmH
2
O
Inspiratory hold time 0.01 s 0 s to 60 s ± 1 % or ± 0.1 s
Expiratory hold time 0.01 s 0 s to 90 s ± 1 % or ± 0.1 s
Peak expiratory flow 0.01 lpm 0 lpm to 300 lpm ± 3 % or ± 2 lpm
Peak inspiratory flow 0.01 lpm 0 lpm to 300 lpm ± 3 % or ± 2 lpm
Lung compliance
1
0.1 ml/
cmH
2
0
0 ml/cmH
2
O to 150 ml/
cmH
2
O
± 5 % or ± 5 ml/cmH
2
O
Flow bias
2
0.01 lpm 0 lpm to 30 lpm ± 2 % or ± 0.5 lpm
1
Inspiratory pause time: > 0.5 s
2
Expiratory pause time: > 0.5 s
Optional accessories
continued
Accessory Kit Parts
1XX0015 Filter, External (bacterial),
1 each
4934FG Adapter, DISS 02 nut and nip-
ple with 1/4 in I.D. hose barb, 1 each
1FT0050 Tubing Adapter, Directional
(15 mm OD x 15 mm OD), 2 each
1FT0049 Tubing Adapter (22 mm OD x
22 mm ID), 2 each
1FT0048 Tubing Adapter (22 mm OD x
22 mm OD), 2 each
1FT0045 Tubing Adapter (15 mm OD x
22 mm OD), 2 each
1FT0046 Tubing Adapter (15 mm OD x
15 mm OD), 2 each
1FT0047 Tubing Adapter (15 mm ID x
15 mm OD), 2 each
1FT0051 Tubing Adapter, Narrow
Bore, 2 each
48478 Barb (luer lock – male to 1/89
in ID tubing), 2 each
1FT0043 Tubing Adapter (1/4 in NPT
male to 1/8 in ID tubing barb fitting),
2 each
1FT0005 Tubing Adapter (luer lock
1/16 in to bulkhead connection), 2 each
2FU0005 Fuse (500 mA)
67535 Tubing 1/8 in 4 ft long, 2 each