Infusion Device Analyzer
Optional accessories
9513-0212 External mini-keyboard,
83-key with PS/2 connector and AT
adapter
71072 Parallel Printer Cable
(D25M-36M)
44277 PCA Trigger/Nurse Call Box
PRINTR/414-US120V Printer, Seiko
DPU-414-30B (120 V power supply)
(additional purchase required: parallel
printer cable, p/n 71072)
PRINTR414-SHK220V Printer, Seiko
DPU-414-30B (220 V power supply)
(additional purchase required: parallel
printer cable, p/n 71072)
61096 Printer (120 V power supply)
61097 Printer (220 V power supply)
3010-0441 Interface Cable, medTester
to IDA 4 Plus (without wedge adapter)
(RS-232; female DB25 to female DB9)
3010-0598 Interface Cable, medTester
to IDA 4 Plus (with or without wedge
adapter) (RS-232; female DB9 to female
DB25)
9513-0221 Barcode Scanner (with
long-reach coil cable with Y connector
for keyboard attachment)
75029 Null Modem Cable (female DB9
to female DB9)
Key features
• Tests up to four infusion pumps
simultaneously
• Compatible with virtually any
type of infusion device
• Instantaneous and average
flow measurement
• Occlusion pressure measure-
ments to 45 psi
• Single- and dual-flow (piggy-
back) testing
• Full PCA testing (bolus volume,
lockout time, and automated
external triggering)
• Autostart mode enables unit to
begin testing only when fluid is
detected
• On-board graphing of pressure
and flow
• Built-in memory to save test
results for printing or down-
loading to computer
• Hydrograph graphical software
to control unit and display
results via PC
• Automated testing through
Fluke Biomedical medTester
5000C (US only)
• RS-232 ports
• Optional keyboard, printer, and
alarm/PCA
IDA 4 Plus Multi-Channel Infusion
Device Analyzer maximizes produc-
tivity with multiple, independent
channels for testing upto four infusion
pumps at once.
The device measures instantaneous
flow, average flow, occlusion pressure,
and analyzes patient-control analgesia
(PCA) units. An optional PCA trigger
box provides automated PCA pump con-
trol, allowing technicians to set
up tests and walk away.
An autostart feature simplifies
syringe pump testing or other tests that
have long startup times.
With built-in memory, the IDA 4 Plus
records test results internally and
provides easy-to-read test-result
graphs right on the analyzer’s screen.
The display is so large numbers can be read from across the room
Additionally, the IDA 4 Plus comes with Hydrograph PC software for cre-
ating full-color graphs and reports. For automated testing, the IDA 4 Plus is
compatible with Fluke Biomedical’s medTester 5000C (US only).
Flow-rate measurement
Technique Calculated by measuring a volume over time
Range 0.5 ml/hr to 1000 ml/hr
Accuracy 1 % of reading ± 1 LSD for flows of 16 ml/hr to 200 ml/hr for
volumes over 20 ml; otherwise, 2 % of reading ± 1 LSD after
delivery of 10 ml
Volume measurement
Technique Volume measured directly by the transducer in minimum
sample sizes of 60 µl
Range 0.06 ml to 9999 ml
Accuracy 1 % of reading ± 1 LSD for flows of 16 ml/hr to 200 ml/hr for
volumes over 20 ml; otherwise, 2 % of reading ± 1 LSD after
delivery of 10 ml
PCA bolus measurement
Technique Volume is measured directly by the transducer in minimum
bolus volumes of 0.5 ml. The measurement is made with a
continuous rate between 0 ml/hr and 30 ml/hr. The bolus flow
rate should be at least four times the basal flow rate for reli-
able detection of boluses
Minimum bolus volume 0.5 ml
Accuracy See volume measurement
Pressure measurement
Technique Direct occlusion of the infusion line and measurement of pres-
sure prior to the glass transducer
Range 0 psi to 45 psi and equivalents in mmHg and kPa
Accuracy 1 % of full scale ± 1 LSD
Back pressure -100 mmHg to 300 mmHg
IDA 4 Plus
Multi-Channel Infusion Device Analyzer
Specications