Infusion Device Analyzer
Optional accessories
9513-0212 External mini-keyboard,
83-key with PS/2 connector and AT
adapter
71072 Parallel Printer Cable
(D25M-36M)
44277 PCA Trigger/Nurse Call Box
PRINTR/414-US120V Printer, Seiko
DPU-414-30B (120 V power supply)
(additional purchase required: parallel
printer cable, p/n 71072)
PRINTR414-SHK220V Printer, Seiko
DPU-414-30B (220 V power supply)
(additional purchase required: parallel
printer cable, p/n 71072)
61096 Printer (120 V power supply)
61097 Printer (220 V power supply)
3010-0441 Interface Cable, medTester
to IDA 4 Plus (without wedge adapter)
(RS-232; female DB25 to female DB9)
3010-0598 Interface Cable, medTester
to IDA 4 Plus (with or without wedge
adapter) (RS-232; female DB9 to female
DB25)
9513-0221 Barcode Scanner (with
long-reach coil cable with Y connector
for keyboard attachment)
75029 Null Modem Cable (female DB9
to female DB9)
Key features
Tests up to four infusion pumps
simultaneously
Compatible with virtually any
type of infusion device
Instantaneous and average
flow measurement
Occlusion pressure measure-
ments to 45 psi
Single- and dual-flow (piggy-
back) testing
Full PCA testing (bolus volume,
lockout time, and automated
external triggering)
Autostart mode enables unit to
begin testing only when fluid is
detected
On-board graphing of pressure
and flow
Built-in memory to save test
results for printing or down-
loading to computer
Hydrograph graphical software
to control unit and display
results via PC
Automated testing through
Fluke Biomedical medTester
5000C (US only)
RS-232 ports
Optional keyboard, printer, and
alarm/PCA
IDA 4 Plus Multi-Channel Infusion
Device Analyzer maximizes produc-
tivity with multiple, independent
channels for testing upto four infusion
pumps at once.
The device measures instantaneous
flow, average flow, occlusion pressure,
and analyzes patient-control analgesia
(PCA) units. An optional PCA trigger
box provides automated PCA pump con-
trol, allowing technicians to set
up tests and walk away.
An autostart feature simplifies
syringe pump testing or other tests that
have long startup times.
With built-in memory, the IDA 4 Plus
records test results internally and
provides easy-to-read test-result
graphs right on the analyzer’s screen.
The display is so large numbers can be read from across the room
Additionally, the IDA 4 Plus comes with Hydrograph PC software for cre-
ating full-color graphs and reports. For automated testing, the IDA 4 Plus is
compatible with Fluke Biomedicals medTester 5000C (US only).
Flow-rate measurement
Technique Calculated by measuring a volume over time
Range 0.5 ml/hr to 1000 ml/hr
Accuracy 1 % of reading ± 1 LSD for flows of 16 ml/hr to 200 ml/hr for
volumes over 20 ml; otherwise, 2 % of reading ± 1 LSD after
delivery of 10 ml
Volume measurement
Technique Volume measured directly by the transducer in minimum
sample sizes of 60 µl
Range 0.06 ml to 9999 ml
Accuracy 1 % of reading ± 1 LSD for flows of 16 ml/hr to 200 ml/hr for
volumes over 20 ml; otherwise, 2 % of reading ± 1 LSD after
delivery of 10 ml
PCA bolus measurement
Technique Volume is measured directly by the transducer in minimum
bolus volumes of 0.5 ml. The measurement is made with a
continuous rate between 0 ml/hr and 30 ml/hr. The bolus flow
rate should be at least four times the basal flow rate for reli-
able detection of boluses
Minimum bolus volume 0.5 ml
Accuracy See volume measurement
Pressure measurement
Technique Direct occlusion of the infusion line and measurement of pres-
sure prior to the glass transducer
Range 0 psi to 45 psi and equivalents in mmHg and kPa
Accuracy 1 % of full scale ± 1 LSD
Back pressure -100 mmHg to 300 mmHg
IDA 4 Plus
Multi-Channel Infusion Device Analyzer
Specications